Incubators are designed to incubate biological indicators which exposed to sterilisation cycle for assurancy of sterilisaiton. Any color change after incubation shows bacterial growth, failure in sterilisation cycle . In each sterilisation cycles; different modules is to be incubated in different tempratures.

Module

Description

Working Temperature, °C

Incubation Time, hrs

1

Incubation for Steam

57 +/- 2

24

2

Incubation for Ethylene Oxide

37 +/- 2

48

3

Incubation for Hydrogen Peroxide

60 +/- 2

24

 

Ref

Description

pc  / Box

Bow Weight

Box Dimension

08-1103

Incubator

10

20

38*38*41